I had had an interesting PD case today that I thought I would share, as this is the
first time I have seen this.
I was sent a patient with a laparoscopically placed right sided double cuffed
coilec PD cather with midline infra-umbilical approach about 5 weeks prior with
intra-Peritoneal fixation. He was sent for complete outflow failure and
inflow with some resistance.
We initially flushed and aspirated a few fibrin plugs from the
catheter - we did not manipulate the PD catheter with wire or
styler. On PD gram I see what I think is encasemrnt of the coil
and I think separation of the catheter with injection. (Video link included)
my suspicion is that the catheter might have torn perhaps by
coming in contact with the staples then completly separated on
contact injection.
the pt will see the surgeon for Laparoendoscopic replacement.
i am interested in knowing if others have seen this and their thoughts.
Raj.